Special MCP-N thermoluminescent detectors (TLD) with better angular dependence / M. Budzanowski, A. JUNG, M. KARCZMARCZYK // Radiation Measurements ; ISSN 1350-4487. — 2024 — vol. 177 art. no. 107245, s. 1–4. — Bibliogr. s. 4, Abstr. — Publikacja dostępna online od: 2024-07-27

Autorzy (3)

Słowa kluczowe

angular dependenceCuPLiF:Mgthermoluminescence

Dane bibliometryczne

ID BaDAP156843
Data dodania do BaDAP2025-02-12
Tekst źródłowyURL
DOI10.1016/j.radmeas.2024.107245
Rok publikacji2024
Typ publikacjiartykuł w czasopiśmie
Otwarty dostęptak
Czasopismo/seriaRadiation Measurements

Abstract

One of the basic parameters in the use of thermoluminescent detectors is the angular characteristics, i.e. receiving different signals depending on the angle of radiation exposure. The TLD detector is typically installed in a slide/badge and here the angular characteristics may have an influence, adding non-uniform coverage of the detectors with filters for the correct determination of the dose as a function of energy, because the energy characteristics are usually not flat. As part of the work, special MCP-N (LiF:Mg, Cu, P) detectors of various shapes were prepared to improve the angular characteristics. The detectors were round pellets with the same circular shape with a diameter of 4.5 mm, and a modified form inside (large and small drilled holes) or a modified surface (cavity or hemisphere). For comparison, standard MCP-N type detectors were used. Detectors in special boxes were exposed to X-rays with an energy of 80 keV. The radiation beam was formed by a medical X-ray apparatus adapted for experiments at the Warsaw University of Technology in Warsaw. Exposures were made for angles of 0, 30, 45, 50, and 90° for a dose of 1.5 mGy. Results show that it is possible to improve the angular characteristics for detectors with modified shapes. Better angular characteristics will allow for more accurate measurements, in particular in comparison tests, for accreditation and other calibrations. Especially improved angular dependence detectors could be useful for H*(10) environmental measurements.

Publikacje, które mogą Cię zainteresować

fragment książki
#152182Data dodania: 6.9.2024
Special thermoluminescent MCP-N (LiF:Mg, Cu, P) detectors with better angular dependence / Maciej Budzanowski, Aleksandra JUNG, Maja Karczmarczyk // W: SSD20 [Dokument elektroniczny] : 20th international conference on Solid State Dosimetry : Viareggio, September 17-22, 2023 ; 8th Solid State Dosimetry Summer School ”Marko Moscovitch” : Pisa, September 13-16, 2023 : [abstracts]. — Wersja do Windows. — Dane tekstowe. — [Pisa : University of Pisa], [2023]. — S. [140]. — Wymagania systemowe: Adobe Reader. — Tryb dostępu: https://docs.google.com/document/d/1UxgiRG_aSEhEQU33FErWiq2q4... [2024-02-28]
artykuł
#132525Data dodania: 17.2.2021
The impact of accidental immersion in selected liquids on the sensitivity and repeatability of MCP-N thermoluminescent detectors / A. JUNG, K. MATUSIAK // Radiation Measurements ; ISSN 1350-4487. — 2021 — vol. 141 art. no. 106525, s. 1–6. — Bibliogr. s. 6, Abstr. — Publikacja dostępna online od: 2021-01-24