Quantitative SPECT/CT imaging of actinium-225 for targeted alpha therapy of glioblastomas / Monika Tulik, Radosław Kuliński, Zbisław TABOR, Beata Brzozowska, Piotr Łaba, Frank Bruchertseifer, Alfred Morgenstern, Leszek Królicki, Jolanta Kunikowska // EJNMMI Physics [Dokument elektroniczny]. - Czasopismo elektroniczne ; ISSN 2197-7364. — 2024 — vol. 11 art. no. 41, s. 1-17. — Wymagania systemowe: Adobe Reader. — Bibliogr. s. 16-17, Abstr. — Publikacja dostępna online od: 2024-05-09

Autorzy (9)

  • Tulik Monika
  • Kuliński Radosław
  • AGHTabor Zbisław
  • Brzozowska Beata
  • Łaba Piotr
  • Bruchertseifer Frank
  • Morgenstern Alfred
  • Królicki Leszek
  • Kunikowska Jolanta

Słowa kluczowe

SPECT/CTActinium-225glioblastomadosimetryquantitative imaging

Dane bibliometryczne

ID BaDAP153355
Data dodania do BaDAP2024-06-07
Tekst źródłowyURL
DOI10.1186/s40658-024-00635-1
Rok publikacji2024
Typ publikacjiartykuł w czasopiśmie
Otwarty dostęptak
Creative Commons
Czasopismo/seriaEJNMMI Physics

Abstract

Background A new, alternative option for patients with recurrent glioblastoma is targeted alpha therapy (TAT), in the form of a local administration of substance P (neurokinin type 1 receptor ligand, NK-1) labelled with 225Ac. The purpose of the study was to confirm the feasibility of quantitative SPECT imaging of 225Ac, in a model reproducing specific conditions of TAT. In particular, to present the SPECT calibration methodology used, as well as the results of validation measurements and their accuracy. Additionally, to discuss the specific problems related to high noise in the presented case. Materials and methods All SPECT/CT scans were conducted using the Symbia T6 equipped with HE collimators, and acquired with multiple energy windows (three main windows: 440 keV, 218 keV, and 78 keV, with three lower scatter energy windows). A Jaszczak phantom with fillable cylindrical sources of various sizes was used to investigate quantitative SPECT/CT imaging characteristics. The planar sensitivity of the camera, an imaging calibration factor, and recovery coefficients were determined. Additionally, the 3D printed model of the glioblastoma tumour was developed and imaged to evaluate the accuracy of the proposed protocol. Results Using the imaging calibration factor and recovery coefficients obtained with the Jaszczak phantom, we were able to quantify the activity in a 3D-printed model of a glioblastoma tumour with uncertainty of no more than 10% and satisfying accuracy. Conclusions It is feasible to perform quantitative 225Ac SPECT/CT imaging. However, there are still many more challenges that should be considered for further research on this topic (among others: accurate determination of ICF in the case of high background noise, better method of background estimation for recovery coefficient calculations, other methods for scatter correction than the dual-energy window scatter-compensation method used in this study).

Publikacje, które mogą Cię zainteresować

artykuł
#164184Data dodania: 8.12.2025
The $[^{99m}Tc]$Tc-HMPAO-labelled white blood cell SPECT/CT as a novel criterion for infective endocarditis diagnosis / Katarzyna Holcman, Andrzej Ząbek, Krzysztof Boczar, Paweł Rubiś, Bogdan ĆMIEL, Wojciech Szot, Agnieszka Stępień, Katarzyna Graczyk, Piotr Podolec, Magdalena Kostkiewicz // International Journal of Cardiology ; ISSN  0167-5273 . — 2024 — vol. 417 art. no. 132545, s. 1–8. — Bibliogr. s. 7–8, Abstr. — Publikacja dostępna online od: 2024-09-10
artykuł
#129878Data dodania: 14.9.2020
The prognostic value of $^{99}mTc$-HMPAO-labeled leucocyte SPECT/CT in cardiac device-related infective endocarditis / Katarzyna Holcman, Paweł Rubiś, Andrzej Ząbek, Bogdan ĆMIEL, Wojciech Szot, Krzysztof Boczar, Sylwia Wiśniowska-Śmiałek, Agnieszka Stępień, Barbara Małecka, Piotr Podolec, Magdalena Kostkiewicz // JACC. Cardiovascular Imaging ; ISSN 1936-878X. — 2020 — vol. 13 iss. 8, s. 1739–1751. — Bibliogr. 1750–1751, Abstr. — Publikacja dostępna online od: 2020-04-15