Szczegóły publikacji

Opis bibliograficzny

Triple-isotope calibration of in-house water standards supplemented by determination of $^{17}O$ content of USGS49-50 reference materials using cavity ring-down laser spectrometry / Anna PIERCHAŁA, Kazimierz RÓŻAŃSKI, Marek DULIŃSKI, Zbigniew GORCZYCA, Robert CZUB // Isotopes in Environmental and Health Studies ; ISSN 1025-6016. — 2021 — vol. 57 no. 3, s. 254–261. — Bibliogr. s. 261, Abstr. — Publikacja dostępna online od: 2021-01-29

Autorzy (5)

Słowa kluczowe

oxygen-18oxygen-17 excesshydrogen-2oxygen-17reference standardscalibration

Dane bibliometryczne

ID BaDAP135395
Data dodania do BaDAP2021-09-07
Tekst źródłowyURL
DOI10.1080/10256016.2021.1875222
Rok publikacji2021
Typ publikacjiartykuł w czasopiśmie
Otwarty dostęptak
Czasopismo/seriaIsotopes in Environmental and Health Studies

Abstract

The procedure of calibrating in-house water standards suitable for routine analyses of triple-isotope composition of water samples using Picarro L2140-i CRDS analyser is presented and discussed. Such standards are indispensable for achieving and maintaining high quality of isotope analyses of water in terms of their precision and accuracy. A set of seven different water standards consisting of three in-house standards and four secondary standards commercially available was calibrated against VSMOW2/SLAP2 primary reference materials. The calibrated standards cover a wide range of isotopic composition, with delta values ranging from close to zero to the values comparable with SLAP2. The apparent consistency of the calibrated values of delta H-2, delta O-18 and d-excess with corresponding certified values for commercially available USGS47-50 standards and the consistency of the calibrated values of delta O-17 and Delta O-17 with its literature values for USGS47-48 standards confirm the high quality of the performed calibration. Moreover, the calibration exercise allowed to obtain delta O-17 and Delta O-17 values for USGS49 and USGS50 standards, not reported so far.

Publikacje, które mogą Cię zainteresować

artykuł
#122466Data dodania: 29.7.2019
High-precision measurements of $\delta^{2}H, \delta^{18}O$ and $\delta^{17}O$ in water with the aid of cavity ring-down laser spectroscopy / Anna PIERCHAŁA, Kazimierz RÓŻAŃSKI, Marek DULIŃSKI, Zbigniew GORCZYCA, Michał MARZEC, Robert CZUB // Isotopes in Environmental and Health Studies ; ISSN 1025-6016. — 2019 — vol. 55 iss. 3, s. 290-307. — Bibliogr. s. 305-307, Abstr. — Publikacja dostępna online od: 2019-04-30
artykuł
#43209Data dodania: 4.2.2009
Isotope effects accompanying evaporation of water from leaky containers / Kazimierz RÓŻAŃSKI, Łukasz CHMURA // Isotopes in Environmental and Health Studies ; ISSN  1025-6016 . — 2008 — vol. 44 no. 1, s. 51–59. — Bibliogr. s. 59. — International workshop on the Isotope effects in evaportation : Pisa, Italy, May 03-05, 2006